 Class II Type A1 BSC Characteristics 
  
  
  
 
						Exhaust approximately 30% of total air handled and recirculate 70%  in cabinet
 
Minimum inflow air velocity is 75 fpm (0.38 m/s) through the front opening
 
May exhaust HEPA filtered air back into the room or through a canopy (to the outdoors)
 
Audiovisual alarm required when connected to a building exhaust system
 
May have contaminated ducts and plenums under positive pressure to the room
 
 
 
	
	  
  
   
  |  |