Class I Biological Safety Cabinet Characteristics



- Minimum intake air velocity of 75 ft/min (0.38 m/s)
- Provides personnel and environmental protection
- Does not provide product protection
- Requires an exhaust blower to pull the air through - usually to the outdoors
- Typical uses of Class I Biological Safety Cabinets include
- Housing centrifuges
- Housing fermenters
- Cage dumping in an animal lab
- Aerating cultures



 
 |
 |